Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2270-2010

Intesys Clinical Suite (ICS) G2 92810 Smart Disclosure. The Smart Disclosure product collects patient...

GE Healthcare / initiated 2008-10-17 / Terminated / root cause: design

Spacelabs Healthcare, Incorporated began a recall of Intesys Clinical Suite (ICS) G2 92810 Smart Disclosure. The Smart Disclosure product collects patient waveforms, alarms, vital signs and 12-lead reports. A... on . The reason FDA records is "Potential for device to print out the wrong waveform data for the time indicated in the printout or display". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2270-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Intesys Clinical Suite (ICS) G2 92810 Smart Disclosure. The Smart Disclosure product collects patient waveforms, alarms, vital signs and 12-lead reports. A 24-hour waveform acquisition is standard, with an upgrade option to 72 hours. The product provides basic trends in both tabular and graphical format, along with a retrospective ECG analysis. The analysis may be defined by patient, and the data may be presented in multiple formats, including a rhythm, a histogram, and a summary view. Standard reports are available from Smart Disclosure, including saved events, disclosure, trends histogram and summary.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Potential for device to print out the wrong waveform data for the time indicated in the printout or display.
Root cause tag
design
FDA root cause
Device Design
Action
On 07/17/2008, Spacelabs Healthcare began sending the MEDICAL DEVICE CORRECTION letter to their consignees. The letter informs the consignees that there is an issue involving ICS G2 Smart Disclosure (Model 92810) which causes the display or printing of an incorrect waveform for the time indicated in the printout or display. The consignees are advised whether to discontinue to use the device until the firm upgrades the system. If the consignees decide to continue to use the device, then they should discontinue any use of the "Refresh" button. When they print or view waveform reports,t hey should utilize the Saved Event view functionality of the product rather than Arrhythmia, Alarm or Waveform Views. If the printing from the Arrhythmia, Alarm, or Waveform views, the consignees must visually inspect the print out and verify that the waveform contains what you expected to see printed. The consignees could contact Spacelabs Healthcare at 1-800-522-7052 or 1-425-657-7200, x5089 for any questions about the corrective actions of the affected device. The firm's field service engineer will make a schedule a mutually agreeable time with the consignees to update the ICS G2 system at free of charge.
Quantity in commerce
135 units
Distribution
Worldwide Distribution -- US, Turkey, Chile, Brazil, Netherlands, Puerto Rico, Switzerland, Australia, Czech Republic, Canada, Italy, Singapore, France, Germany, and Israel.
Product codes and lots
Software version V04.00.02 was shipped on 06/30/2008.  The 92810 is a software application installed on an off-the-shelf computer and is not serialized. Spacelabs uses software versions and ship dates for identification and traceability purposes.