Tornier AFFINITI Humeral Stem Standard 14mm, Catalog No. 0020014. Sterile. Manufactured for Tornier, 10750...
Tournier, Inc. began a recall of Tornier AFFINITI Humeral Stem Standard 14mm, Catalog No. 0020014. Sterile. Manufactured for Tornier, 10750 Cash Rd., Stafford, TX 77477. Tornier's Affiniti... on . The reason FDA records is "Tornier, Medical has received a report of a mislabeled sterile AFFINITI head product. As a result, Tornier has decided to voluntarily recall the potentially affected lots, including two (2) lots of AFFINITI Stems". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0819-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tournier, Inc.
- Product
- Tornier AFFINITI Humeral Stem Standard 14mm, Catalog No. 0020014. Sterile. Manufactured for Tornier, 10750 Cash Rd., Stafford, TX 77477. Tornier's Affiniti Shoulder System is a non-constrained glenohumeral prosthesis intended for use as a total or hemi-shoulder replacement system. The total shoulder consists of a metal humeral stem, a metal humeral head and an ultrahigh molecular weight polyethylene glenoid. The hemi-shoulder consists of a metal humeral stem and a metal humeral head.
- Product code
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Reason for recall
- Tornier, Medical has received a report of a mislabeled sterile AFFINITI head product. As a result, Tornier has decided to voluntarily recall the potentially affected lots, including two (2) lots of AFFINITI Stems.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- An Urgent Product Recall letter, dated 10/17/08, was hand delivered to end users by distributors and sales representatives beginning 10/20/08. The recall letter identified the issue and requested end users to immediately discontinue use of the affected products and return any unused affected products to Tornier. Distributors were to notify any and all accounts that may be end users or to provide this information to Tornier, whom will contact the end users. A questionnaire and return instructions were provided with the letter. The questionnaire was asked to be completed and returned to Tornier as soon as possible. For additional information, contact Tornier Customer Service at 1-888-494-7950.
- Quantity in commerce
- 401 total
- Distribution
- AR, CA, FL, NC, NE, OH, PA, TN, TX, WA, and WI
- Product codes and lots
- Lot No.: AA1434
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.