Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0824-2009

Tornier AFFINITI Humeral Head Eccentric 56mm x 21mm, Catalog No. 0020038. Sterile. Manufactured for Tornier...

Tournier, Inc. / initiated 2008-10-20 / Terminated / root cause: labeling and UDI

Tournier, Inc. began a recall of Tornier AFFINITI Humeral Head Eccentric 56mm x 21mm, Catalog No. 0020038. Sterile. Manufactured for Tornier, 10750 Cash Rd., Stafford, TX 77477. Tornier's... on . The reason FDA records is "Tornier, Medical has received a report of a mislabeled sterile AFFINITI head product. As a result, Tornier has decided to voluntarily recall the potentially affected lots, including two (2) lots of AFFINITI Stems". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0824-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tournier, Inc.
Product
Tornier AFFINITI Humeral Head Eccentric 56mm x 21mm, Catalog No. 0020038. Sterile. Manufactured for Tornier, 10750 Cash Rd., Stafford, TX 77477. Tornier's Affiniti Shoulder System is a non-constrained glenohumeral prosthesis intended for use as a total or hemi-shoulder replacement system. The total shoulder consists of a metal humeral stem, a metal humeral head and an ultrahigh molecular weight polyethylene glenoid. The hemi-shoulder consists of a metal humeral stem and a metal humeral head.
Product code
KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason for recall
Tornier, Medical has received a report of a mislabeled sterile AFFINITI head product. As a result, Tornier has decided to voluntarily recall the potentially affected lots, including two (2) lots of AFFINITI Stems.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
An Urgent Product Recall letter, dated 10/17/08, was hand delivered to end users by distributors and sales representatives beginning 10/20/08. The recall letter identified the issue and requested end users to immediately discontinue use of the affected products and return any unused affected products to Tornier. Distributors were to notify any and all accounts that may be end users or to provide this information to Tornier, whom will contact the end users. A questionnaire and return instructions were provided with the letter. The questionnaire was asked to be completed and returned to Tornier as soon as possible. For additional information, contact Tornier Customer Service at 1-888-494-7950.
Quantity in commerce
401 total
Distribution
AR, CA, FL, NC, NE, OH, PA, TN, TX, WA, and WI
Product codes and lots
Lot No.: AA1307