Z-1070-2009
OneTouch SureStep Test Strips, Part numbers 020-052-01 and 010-359-04, for use with SureStep brand Blood...
Lifescan, Inc. / initiated 2008-11-25 / Terminated / root cause: manufacturing process
Lifescan Inc began a recall of OneTouch SureStep Test Strips, Part numbers 020-052-01 and 010-359-04, for use with SureStep brand Blood Glucose Meters, manufactured by LifeScan Inc... on . The reason FDA records is "Deformity in the test strip may result in insufficient blood transfer to reaction area, resulting in inaccurate test results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1070-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lifescan, Inc.
- Product
- OneTouch SureStep Test Strips, Part numbers 020-052-01 and 010-359-04, for use with SureStep brand Blood Glucose Meters, manufactured by LifeScan Inc., Milpitas, CA
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- Deformity in the test strip may result in insufficient blood transfer to reaction area, resulting in inaccurate test results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Urgent: Medical Device Recall letters, dated 11/24/2008 were issued to all consignees on November 25, 2008, requesting return of affected lots to LifeScan and instructing how to obtain replacement strips free of charge. A request for sub-recall communications were made as well, and the firm posted notification on its website. Customer service and Sales account reps were provided information to address questions regarding the recall.
- Quantity in commerce
- 23,288 vials
- Distribution
- Worldwide Distribution -- USA and the Caribbean.
- Product codes and lots
- Lot codes 2802961 and 2802962