Z-0858-2009

RF Ablation System Foot Switch. (Can be used with Boston Scientific Maestro 3000 and EPT-1000XP Cardiac...

Boston Scientific Corporation / initiated 2008-11-25 / Terminated / root cause: manufacturing process

Boston Scientific Corp began a recall of RF Ablation System Foot Switch. (Can be used with Boston Scientific Maestro 3000 and EPT-1000XP Cardiac Ablation Cardiac Ablation Systems). Manufactured by... on . The reason FDA records is "Product malfunction may result in delivery of energy without depressing the foot switch that controls delivery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0858-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
RF Ablation System Foot Switch. (Can be used with Boston Scientific Maestro 3000 and EPT-1000XP Cardiac Ablation Cardiac Ablation Systems). Manufactured by Altech Corporation, Flemmington NJ and distributed by Boston Scientific Corporation, San Jose, CA, catalog numbers/UPN numbers 21840/M004 218400. Indicated for use in cardiac ablation procedures.
Product code
LPB Cardiac Ablation Percutaneous Catheter
Reason for recall
Product malfunction may result in delivery of energy without depressing the foot switch that controls delivery.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Recall initiated on November 25, 2008. Urgent Company Initiated Medical Device Recall - Immediate Action Required letters, dated 11/26/2008, were sent to consignees via Federal Express and included return and reply forms for consignees to respond. The letters asked that further distribution or use of any foot switch affected by this recall be ceased immediately. Attached to the letter was a document entitled "Instructions for Determining Affected Product," which provided instructions and serial numbers for determining if customers have product affected by this recall. In addition, a document entitled "Recall Instructions" was also attached. Affected product should be segregated and returned to Boston Scientific in accordance to the "Recall Instructions," and the Reply and Verification Tracking Form should be completed and returned. Customers will receive replacement product for all recalled product returned to Boston Scientific. Local Sales Reps should be contacted in regards to any questions customers may have pertaining to this recall.
Quantity in commerce
322 units
Distribution
Worldwide Distribution --- including USA and countries of France, Great Britain, Saudi Arabia, Spain, Canada, Portugal, Germany, Italy, Australia, Russia, Netherlands, Ireland, Libya.
Product codes and lots
M004 218400, Serial numbers starting in 0430200, 0530100, 0606100, 0610000, 0620200, 0623700, 0626500, 0629900, 0634800, and ending in 001-405.