Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0537-2010

Otto Bock Modular Knee Joint, 3R66, Otto Bock HealthCare GmbH, Max-Nader-StraBe 15 . 37115 Duderstadt...

Bock,Otto,Orthopedic Ind,Inc / initiated 2008-10-28 / Terminated / root cause: manufacturing process

Bock,Otto,Orthopedic Ind,Inc began a recall of Otto Bock Modular Knee Joint, 3R66, Otto Bock HealthCare GmbH, Max-Nader-StraBe 15 . 37115 Duderstadt /Germany. Pediatric Prosthetic Device used by Lower Limb... on . The reason FDA records is "Otto Bock Health Care LP has discovered that in some cases, the securing screws of the joint axes may not have been properly mounted. This can cause the axes to gradually move out of the joint and result in a risk of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0537-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bock,Otto,Orthopedic Ind,Inc
Product
Otto Bock Modular Knee Joint, 3R66, Otto Bock HealthCare GmbH, Max-Nader-StraBe 15 . 37115 Duderstadt /Germany. Pediatric Prosthetic Device used by Lower Limb amputees.
Product code
ISY Joint, Knee, External Limb Component
Reason for recall
Otto Bock Health Care LP has discovered that in some cases, the securing screws of the joint axes may not have been properly mounted. This can cause the axes to gradually move out of the joint and result in a risk of falling for the user.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Consignees were sent on 10/28/08 a Otto Bock "Urgent: Correction Notice" letter dated October 27, 2008. The letter described the product and problem. It provided instructions if the product was in stock and instructed the consignee to contact the patients (a patient letter was provided) and to complete and return the reply card. Consignees were to contact Otto Bock at their Customer Service to request a return authorization.
Quantity in commerce
9
Distribution
The product was only distributed within the US by the US company. Any OUS distribution was handled by the manufacturer in Germany.
Product codes and lots
Serial numbers: 200803001, 200803002, 200803003, 200803004, 200803005, 200803006, 200803007, 200803008, and 200803009.