ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; manufactured by Inpecco SpA, Segrate...
Abbott Laboratories, Inc began a recall of ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; manufactured by Inpecco SpA, Segrate (MI), Italy on . The reason FDA records is "When an ARCHITECT c16000 System is connected to an ACCELERATOR APS System, the software on the ACCELERATOR APS System may not generate an error warning when a Sample Presentation/Sample Queue error occurs, which could...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2221-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; manufactured by Inpecco SpA, Segrate (MI), Italy.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- When an ARCHITECT c16000 System is connected to an ACCELERATOR APS System, the software on the ACCELERATOR APS System may not generate an error warning when a Sample Presentation/Sample Queue error occurs, which could result in contamination of subsequent samples.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Abbott Laboratories notified Customers of the affected device via a "Product Correction" letter dated October 27, 2008. Consignees provided with a temporary work around until a service representative upgrades the system software. For further information, contact your local Abbott Laboratories, Inc. service representative.
- Quantity in commerce
- 15 units
- Distribution
- International Distribution -- Spain, Italy, the United Kingdom, Sweden, Switzerland, the Netherlands, France, Canada and Scotland.
- Product codes and lots
- Serial Numbers: ABT 042, ABT 055, ABT 037, ABT 016, ABT 041, ABT 046, ABT 035, ABT 040, ABT 039, ABT 033, ABT 028, ABT 011, ABT 006 and ABT 007.