Z-0862-2009
Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in...
Miltex, Inc. / initiated 2008-11-17 / Terminated / root cause: manufacturing process
Miltex, Inc. began a recall of Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades on . The reason FDA records is "Blade may break: Over milling of the blade resulted in tip more narrow than specified". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0862-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Miltex, Inc.
- Product
- Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.
- Product code
- GES Blade, Scalpel
- Reason for recall
- Blade may break: Over milling of the blade resulted in tip more narrow than specified.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- An Urgent Medical Device Recall letter dated 11/17/08 with an accompanying fax back form was sent to each customer. Each distributor was instructed to review their inventory, complete the form, and return any affected product. The distributors were also instructed to conduct a sub recall notification.
- Quantity in commerce
- 200 boxes
- Distribution
- Nationwide Distribution and Canada.
- Product codes and lots
- Lot number S07C06