Z-0647-2009
CT syngo Dual Energy Viewer on the syngo Multimodality Work Place (MMWP) and on the CT workplace, Model...
Siemens Medical Solutions USA, Inc. / initiated 2008-11-13 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of CT syngo Dual Energy Viewer on the syngo Multimodality Work Place (MMWP) and on the CT workplace, Model number 10094833. Picture Archiving and Communications... on . The reason FDA records is "Indicated orientation does not match actual orientation of the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0647-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- CT syngo Dual Energy Viewer on the syngo Multimodality Work Place (MMWP) and on the CT workplace, Model number 10094833. Picture Archiving and Communications System.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Indicated orientation does not match actual orientation of the patient.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued a Customer Safety Advisory dated 10/8/08 to affected customers via Update Instructions CT052/08/S. The letter informs customers of the potential issue and provides instructions to avoid its occurrence. Customers should consider "no saving of calculated Dual Energy data sets from non-supine and mirror-option scans via "Save Sets CT"-button." The firm also highly recommends for customers "to check the correct display of "left", "anterior" and "posterior"markings, e.g. by checking the position of the spine with respect to the posterior marking in the image." The error should be fixed within certain upgrade and installation activities on the affected software.
- Quantity in commerce
- 374 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Software versions: syngo MMW 2007C, syngo MMWP 2008A, syngo MMWP 2008B, syngo MMWP 2008C, syngo CT 2007A, and syngo CT 2008G.