Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0644-2009

Edwards Lifesciences CardioVations, EndoClamp aortic catheter 100 cm, REF EC1001, Sterile R, individually...

Edwards Lifesciences LLC / initiated 2008-11-26 / Terminated / root cause: manufacturing process

Edwards Lifesciences Research Medical, Inc. began a recall of Edwards Lifesciences CardioVations, EndoClamp aortic catheter 100 cm, REF EC1001, Sterile R, individually packaged, Edwards Lifesciences LLC, Irvine, CA 92614... on . The reason FDA records is "Balloon rupture: Catheter balloon may rupture during cardiopulmonary bypass procedure. For further information, please telephone Edwards Lifesciences Research Medical, Inc. at 801-565-6173". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0644-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards Lifesciences CardioVations, EndoClamp aortic catheter 100 cm, REF EC1001, Sterile R, individually packaged, Edwards Lifesciences LLC, Irvine, CA 92614. This product is indicated for use in patients undergoing cardiopulmonary bypass. It occludes the ascending aorta, vents the aortic root and allows delivery of cardioplegia solution to arrest the heart. The pressure lumen allows monitoring of the aortic root pressure.
Product code
DXC Clamp, Vascular
Reason for recall
Balloon rupture: Catheter balloon may rupture during cardiopulmonary bypass procedure. For further information, please telephone Edwards Lifesciences Research Medical, Inc. at 801-565-6173.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
All consignees were notified by letter on 11/26/2008 and requested to return any unused product to Edwards. The letter was entitled "PRODUCT RECALL." Questions could be referred to the Edwards representative or call 800-424-3278.
Quantity in commerce
1267 units
Distribution
Globally - USA including the states of AK, AL, CA, CO, CT, DC, DE, FL, GA, IA, IL, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, and WV. VA facilities in VA and WI. No military distribution. Foreign distribution to Australia, Canada, Switzerland and UK.
Product codes and lots
Lot numbers: MS0408007, MS0508033, MS0508040, 561241, 562406, 567576, 569133, 569134, 569135, 575695, 579683, and 581317.