Z-0848-2009
Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/22 cm, sterile, Cook Urological, Inc., Spencer...
Cook Incorporated / initiated 2008-07-31 / Terminated / root cause: manufacturing process
Vance Products Incorporated began a recall of Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/22 cm, sterile, Cook Urological, Inc., Spencer, IN; Ref EUSCFS-070022-14-13; GPN REF G18133... on . The reason FDA records is "The bond may break where the two materials meet, resulting in separation of the stent". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0848-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Cook Urological Endoureterotomy Stent Set, 14Fr/13 cm, 7.0Fr/22 cm, sterile, Cook Urological, Inc., Spencer, IN; Ref EUSCFS-070022-14-13; GPN REF G18133. Ureteral stent.
- Product code
- FAD Stent, Ureteral
- Reason for recall
- The bond may break where the two materials meet, resulting in separation of the stent.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified via "Urgent: Recall Notice" letter dated 7/31/08. The letter described the failure, instructed that the products be isolated and returned to the firm. For additional information, contact Cook Urological at 1-812-339-2235, ext. 7241.
- Quantity in commerce
- 1739 of all products.
- Distribution
- Nationwide, Australia, Canada, Colombia, Hong Kong, India, Ireland, Israel, Japan, Malaysia, Mexico, South Korea, Taiwan, Venezuela and Viet Nam
- Product codes and lots
- Lots UF1591697, UF1537682, UF1562307, UF1420730, UF1404622, UF1402891, UF1531085, UF1526725, UF1396538, UF1406357, UF1533678, UF1542186, UF1570187, UF1575838, UF1586147, UF1581083, UF1555888 and UF1389149.