Z-0927-2009

Artiste MV SA Linear Accelerator with syngo RT Therapist Version 4.1 build 67 and build 122. Manufactured by...

Siemens Medical Solutions USA, Inc. / initiated 2008-10-23 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Artiste MV SA Linear Accelerator with syngo RT Therapist Version 4.1 build 67 and build 122. Manufactured by Siemens Medical Solutions Oncology Care Systems... on . The reason FDA records is "If a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0927-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artiste MV SA Linear Accelerator with syngo RT Therapist Version 4.1 build 67 and build 122. Manufactured by Siemens Medical Solutions Oncology Care Systems, Concord, CA.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
If a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the immediate resumption feature, syngo RT Therapist Express systems will display and record a second delivery of a prior beam and not the delivery of the current beam in certain circumstanc
Root cause tag
software
FDA root cause
Software design
Action
An Urgent Medical Device Customer Safety Letter dated October 23, 2008 was issued to consignees via certified mail. The letter identified the issue and preventive measures for consignees. Monitoring will be done via return cards. Follow-up to be conducted via telephone.
Quantity in commerce
36 units
Distribution
Worldwide Distribution including USA and countries of Netherlands, Germany, Poland, Norway, Australia, Hungary, Belgium, New Zealand, France, Malaysia, Spain, and Canada.
Product codes and lots
Part Number 8139789, Serial Numbers: 5190, 5350, 5355, 5364, 5367, 5374, 5377, 5398, 5405, 5410, 5419, 5424, 5165, 5166, 5202, 5238, 5298, 5306, 5342, 5352, 5365, 5366, 5373, 5378, 5379, 5380, 5397, 5399, 5400, 5401, 5402, 5409, 5414, 5416, 5423 and 5425.