LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit...
Millipore Corporation began a recall of LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit Catalog #3121 Simulfluor Flu A / Flu B DFA kit on . The reason FDA records is "Certain lots of Simulfluor Flu A / Flu B Reagent may cross-react with Respiratory Synctial Virus. The continued use of Simulfluor Flu A / Flu B Reagent Catalog #5250 may lead to a potential misdiagnosis of influenza B". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0954-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- EMD Millipore Corporation
- Product
- LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit Catalog #3121 Simulfluor Flu A / Flu B DFA kit.
- Product code
- GNW Antisera, Cf, Influenza Virus A, B, C
- Reason for recall
- Certain lots of Simulfluor Flu A / Flu B Reagent may cross-react with Respiratory Synctial Virus. The continued use of Simulfluor Flu A / Flu B Reagent Catalog #5250 may lead to a potential misdiagnosis of influenza B.
- FDA root cause
- Pending
- Action
- Millipore is performing a voluntary recall by notifying customers and distributors by telephone on November 24, 2008. In addition, an Urgent: Medical Device Recall notification letter with the same date was sent via certified mail. Customers were advised to examine their stock immediately and discontinue the use of the specified lots. Customers were instructed to complete the enclosed Effective Check Survey confirming disposal of defective lots and return this form to Millipore's Regulatory Department. Customer service is to be contacted at 1-800-437-7500 for replacement product. Millipore Corporation will issue credit or replace the recalled lots that customers currently have in inventory.
- Quantity in commerce
- 258 units
- Distribution
- Worldwide Distribution -- USA including states of AL, CA, CO, CT, FL, ID, IL, KS, MA, MI, MS, NM, NY, OH, OK, OR, PA, SD, TX, UT, VA, VT, WA, WV, and WI and countries of Canada, Mexico, Italy, Belgium,Germany, France, Spain, and United Kingdom.
- Product codes and lots
- Lot Numbers JBC1531414, TR1458084, TR1480746, JBC1531421, JH1534500, TR1501923, TR1417106, and TR1422506.