Z-0718-2009

King Systems Universal Flex2 Breathing Circuit, Latex Free, 50 per case, King Systems Corporation...

King Systems Corporation. / initiated 2008-08-01 / Terminated

King Systems Corp. began a recall of King Systems Universal Flex2 Breathing Circuit, Latex Free, 50 per case, King Systems Corporation, Noblesville, IN; REF # DF375-6121Z. This breathing circuit... on . The reason FDA records is "The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0718-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
King Systems Corporation.
Product
King Systems Universal Flex2 Breathing Circuit, Latex Free, 50 per case, King Systems Corporation, Noblesville, IN; REF # DF375-6121Z. This breathing circuit is a device intended to administer medical gases and/or anesthetic gases to a patient during anesthesia for inhalation, or medical gases to a patient during respiratory care inhalation.
Product code
CAI Circuit, Breathing (W Connector, Adaptor, Y Piece)
Reason for recall
The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.
FDA root cause
Pending
Action
Consignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand.
Quantity in commerce
Approx. 500,000 units of all products.
Distribution
Worldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru.
Product codes and lots
Lots IKB40, IKCV6, IKG74, IKI88, IKIZ1, IKLW5, IKP70, IKRK2, IKTV7, IKW65, IKXH8, IKYL5, IL0A8, IL2T1, IL4L1, IL4L2, IL4L3, IL7J1, IL8X1 and ILBN7.