Z-0931-2009

Low Profile U-Joint Driver for the SynFix Mini-Open System, Catalog number 03.802.231. The Synthes SynFix LR...

Synthes GmbH / initiated 2008-11-20 / Terminated / root cause: design

Synthes Spine began a recall of Low Profile U-Joint Driver for the SynFix Mini-Open System, Catalog number 03.802.231. The Synthes SynFix LR is indicated for use in patients with degenerative... on . The reason FDA records is "screwdriver tip may break off and lodge in the screw head". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0931-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Low Profile U-Joint Driver for the SynFix Mini-Open System, Catalog number 03.802.231. The Synthes SynFix LR is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the SynFix LR can be packed with autograft.
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
screwdriver tip may break off and lodge in the screw head
Root cause tag
design
FDA root cause
Device Design
Action
The recalling firm emailed all Spine Sales Consultants on 11/20/08. They were instructed to immediately examine their stock and discuss appropriate technique with surgeon users. In addition, a Medical Device Correction letter dated 11/19/08 was issued via USPS certified mail to the medical facilities. This initial letter stated the issue and requested the users take the following actions: discard all SynFix Technique Guides and replace with the updated literature contained in the mailing packet with the letter, and is identified by code J7022-E (Rev. E); to ensure receipt and understanding of the information, the Verification Section of the letter needs to be completed and returned to Synthes by either fax: 610-719-5120 or scan/e-mail: [email protected]. An Urgent: Medical Device Recall letter, dated 1/16/2009, was also sent out. This latter letter requested that users examine their stock immediately and remove any of the affected devices from their inventory. Customers are also to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number; complete the Verification Section; and return the letter with the recalled product to the address listed. There are also actions for those who do not have the affected product. Questions should be directed to 800-620-7025 x 5453 or 610-719-5453.
Quantity in commerce
381 units
Distribution
Class II Recall - Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA ,SC, TN, TX, UT, VA, WA, and WI, and countries of Canada and Switzerland.
Product codes and lots
Lot numbers: 2317363, 2331806, 2331810, 2387975, 2387976, 2401982, 5637286, 5644605, 5660377, 5664629, 5826243, and 5826244.