Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650...

Stryker Corp. / initiated 2008-12-01 / Terminated

Stryker Endoscopy began a recall of Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801... on . The reason FDA records is "Leak: Leak-resistant barrier may not properly seal and leak when an instrument is inserted, and keep leaking after instrument is removed". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0868-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows instruments and scopes access to the joint while maintaining a seal against the loss of saline fluid used to keep the joint clean. The seal is formed by two elastomeric valves. The first controls loss of water when there is no instrument or scope therein, while the second provides a seal when the cannula is in use. The valves operate automatically. Manufactured by Stryker Endoscopy Puerto Rico
Product code
HRX Arthroscope
Reason for recall
Leak: Leak-resistant barrier may not properly seal and leak when an instrument is inserted, and keep leaking after instrument is removed.
FDA root cause
Component change control
Action
Recall initiated 12/01/2008, Sales representatives were notified via telephone and e-mail. Product advisories entitled ""Urgent: Device Removal" and dated December 1, 2008, were sent to all consignees. All unused product is to be returned to the firm. For further information, please contact Stryker Endoscopy by telephone at 408-754-2124.
Quantity in commerce
4,194 boxes (24,750 units)
Distribution
Worldwide Distribution --- including USA and countries of Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, New Zealand, Poland, Portugal, Sweden, Spain, Singapore, Switzerland, and UK.
Product codes and lots
Lot numbers 08135AG2 through 08294AG2.