Z-1395-2009
Bio-Rad Variant II Hemoglobin Testing System with CDM v 4.0, model number 270-2000, Manufactured by Bio-Rad...
Bio-Rad Laboratories Inc. / initiated 2008-12-09 / Terminated / root cause: labeling and UDI
Bio-Rad Laboratories Inc began a recall of Bio-Rad Variant II Hemoglobin Testing System with CDM v 4.0, model number 270-2000, Manufactured by Bio-Rad Laboratories, Hercules, CA Measurement of percent... on . The reason FDA records is "The wrong barcode number and tube position number may be assigned to samples leading to discrepant sample results". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1395-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- Bio-Rad Variant II Hemoglobin Testing System with CDM v 4.0, model number 270-2000, Manufactured by Bio-Rad Laboratories, Hercules, CA Measurement of percent hemoglobin Alc is effective in monitoring long-term glucose control in individuals with diabetes mellitus.
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- The wrong barcode number and tube position number may be assigned to samples leading to discrepant sample results.
- Root cause tag
- labeling and UDI
- FDA root cause
- Pending
- Action
- Recall initiated on December 10, 2008. Notification conducted by phone call and fax for domestic consignees. Each subsidiary will receive the subsidiary Urgent: Medical Device Correction notice, customer medical device correction response form and number of customers distributed to by e-mail. Each subsidiary must contact customers and provide the customer notice and response form as needed. The notice stated that Bio-Rad Laboratories would release a follow up communication providing a permanent solution upon completion of an investigation. In the interim, it was recommended that customers carefully review the Summary Report for duplicate barcode IDs or or tube position numbers within the same rack. If there are any questions or assistance needed, the local regional Bio-Rad office should be contacted.
- Quantity in commerce
- 42 units
- Distribution
- Worldwide Distribution -- USA, Argentina, Australia, Brazil, Canada, China, Czech Republic, France, Germany, Hungary, India, Italy, Mexico, New Zealand, Singapore, South Korea, Sweden, Thailand, and the UK.
- Product codes and lots
- CDM version 4.0 software