Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0944-2009

Symbiq One-Channel list number 16026-04 and Two-Channel list number 16027-04 Infusion System, a general...

Abbott Medical / initiated 2008-12-03 / Terminated / root cause: software

Hospira Inc began a recall of Symbiq One-Channel list number 16026-04 and Two-Channel list number 16027-04 Infusion System, a general purpose infuser, Distributed and Manufactured by... on . The reason FDA records is "Hypersensitivity to normal motor friction/drag may cause malfunction errors and alarms that are designed to stop pump operations". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0944-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Symbiq One-Channel list number 16026-04 and Two-Channel list number 16027-04 Infusion System, a general purpose infuser, Distributed and Manufactured by Hospira Inc, Morgan Hill, CA. Is a general purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural, or irrigation routes of administration. Intended primarily for use in the hospital setting and can be used in other acute and non-acute care areas, such as but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient/Surgical Centers, Long Term Care, Urgent Care, Transport and Physician Offices.
Product code
FRN Pump, Infusion
Reason for recall
Hypersensitivity to normal motor friction/drag may cause malfunction errors and alarms that are designed to stop pump operations.
Root cause tag
software
FDA root cause
Software design
Action
Recall initiated 12/03/2008. Consignees were notified of the issue via an Urgent: Device Field Correction letter, dated 12/3/2008, and were requested to respond appropriately to any alarm or warnings displayed by the device. The firm is making arrangements to screen each Symbiq pump in customer facilities and implement the appropriate enhancements.
Quantity in commerce
combined 11,840 units
Distribution
Worldwide Distribution --- including USA and countries of Canada and Australia.
Product codes and lots
All products manufactured between 12/19/06 to 12/02/08 with list numbers 16026-04-01/03/04, 51/52, 79/80 and 81/82, or 16027-04-01/03/04, 51/52, 53/54, 79/80, and 81/82.