Z-0885-2009
Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional - Articular surface, Size Blue, 14...
Zimmer GmbH / initiated 2008-12-12 / Terminated / root cause: manufacturing process
Zimmer Inc. began a recall of Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional - Articular surface, Size Blue, 14 MM Height, nonsterile, Zimmer, Warsaw, IN; REF... on . The reason FDA records is "Some 12 mm thick devices may have been etched and packaged as 14 mm thick, and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0885-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional - Articular surface, Size Blue, 14 MM Height, nonsterile, Zimmer, Warsaw, IN; REF 5971-50-14. Used during total knee arthroplasty surgery to determine the required thickness of the tibial bearing surface to be implanted. Reusable, non-implantable.
- Product code
- HWT Template
- Reason for recall
- Some 12 mm thick devices may have been etched and packaged as 14 mm thick, and vice versa.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Zimmer sales force and distribution sites were notified of the recall via email on 12/15/08 and instructed to inventory, return the affected products and to provide a copy of the recall letter to those customers to whom they had further distributed the products. The 'Urgent: Device Recall" letter dated 12/12/08, described the problem and the potential risks. For additional information, contact Zimmer, Inc., at 1-800-613-6131.
- Quantity in commerce
- 131 of both products.
- Distribution
- Nationwide, Canada, Denmark, Finland, Germany, Poland, Spain, Sweden and United Kingdom.
- Product codes and lots
- Lots 60969810 and 60969811.