Z-0886-2009

Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional - Articular surface, Size Blue, 12...

Zimmer GmbH / initiated 2008-12-12 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional - Articular surface, Size Blue, 12 MM Height, nonsterile, Zimmer, Warsaw, IN; REF... on . The reason FDA records is "Some 12 mm thick devices may have been etched and packaged as 14 mm thick, and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0886-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer NEXGEN Complete Knee Solution Cruciate Retaining (CR) Provisional - Articular surface, Size Blue, 12 MM Height, nonsterile, Zimmer, Warsaw, IN; REF 5971-50-12. Used during total knee arthroplasty surgery to determine the required thickness of the tibial bearing surface to be implanted. Reusable, non-implantable.
Product code
HWT Template
Reason for recall
Some 12 mm thick devices may have been etched and packaged as 14 mm thick, and vice versa.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Zimmer sales force and distribution sites were notified of the recall via email on 12/15/08 and instructed to inventory, return the affected products and to provide a copy of the recall letter to those customers to whom they had further distributed the products. The 'Urgent: Device Recall" letter dated 12/12/08, described the problem and the potential risks. For additional information, contact Zimmer, Inc., at 1-800-613-6131.
Quantity in commerce
131 of both products.
Distribution
Nationwide, Canada, Denmark, Finland, Germany, Poland, Spain, Sweden and United Kingdom.
Product codes and lots
Lots 61007188, 60987590, 60969807, 60976981 and 60976982.