Z-0976-2009

Turbo Elite 0.9mm Turbo Elite (TE) RX, Model #410-154, tip diameter 0.9mm/.038 in, 4 Fr, sterile EO...

Spectranetics Corp. / initiated 2008-07-04 / Terminated / root cause: labeling and UDI

Spectranetics Corporation began a recall of Turbo Elite 0.9mm Turbo Elite (TE) RX, Model #410-154, tip diameter 0.9mm/.038 in, 4 Fr, sterile EO. Manufactured by Spectranetics, 96 Talamine Court, Colorado... on . The reason FDA records is "The catheter pouch label was incorrectly labeled with the wrong size". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0976-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Spectranetics Corp.
Product
Turbo Elite 0.9mm Turbo Elite (TE) RX, Model #410-154, tip diameter 0.9mm/.038 in, 4 Fr, sterile EO. Manufactured by Spectranetics, 96 Talamine Court, Colorado Springs, CO 80907. The pouch label of the 1.4mm TE Over the Wire (OTW) catheter, Ref # 414-151, L/N# 080417D was incorrectly labeled for size as 0.9mm TE RX, Ref# 410-154, L/N#080417D. The box labels are correct and match the devices in the pouch. The Spectranetics TURBO-BoosterTM guiding catheter is designed for directing and supporting Spectranetics laser catheters for use in the treatment of infrainguinal stenoses and occlusions.
Product code
MJQ
Reason for recall
The catheter pouch label was incorrectly labeled with the wrong size.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
The mislabeled catheter was returned to Spectranetics. The second possible mislabeled catheter was visually inspected by a Spectranetics representative and confirmed the catheter was labeled correctly.
Quantity in commerce
2 units
Distribution
Nationwide Distribution including states of TX and FL.
Product codes and lots
Lot Number: 080417D.