Z-0991-2009
Zimmer Minimally Invasive Solutions Technology Offset Rasp Handle, Right, 30 degrees, Zimmer, Warsaw, IN; REF...
Zimmer GmbH / initiated 2008-12-19 / Terminated / root cause: labeling and UDI
Zimmer Inc. began a recall of Zimmer Minimally Invasive Solutions Technology Offset Rasp Handle, Right, 30 degrees, Zimmer, Warsaw, IN; REF 00-7806-035-02. The offset rasp handle is a... on . The reason FDA records is "The c-clip located near the connection end may fracture or fall off during use in surgical procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0991-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Minimally Invasive Solutions Technology Offset Rasp Handle, Right, 30 degrees, Zimmer, Warsaw, IN; REF 00-7806-035-02. The offset rasp handle is a surgical instrument used with various sizes of femoral rasps to prepare the femoral canal for a hip implant.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The c-clip located near the connection end may fracture or fall off during use in surgical procedures.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- U.S. consignees were notified by "Urgent: Device Recall" letter issued via e-mail on December 19, 2008. The letter described the problem and instructed customers to immediately quarantine and return the recalled product along with the completed Inventory Return Certification form to the Zimmer, Inc. Distribution Center. Consignees were also instructed to fax a copy of the completed Inventory Return Certification Form at 1-574-371-8603. International consignees were also notified, but Zimmer, Inc. will gather the instruments from those customers. Direct questions about the recall to Zimmer, Inc. at 1-800-613-6131.
- Quantity in commerce
- 1,461 of both products.
- Distribution
- Worldwide Distribution --- including USA, Belgium, Czech Republic, Finland, Germany, Greece, Italy, Poland, Romania, Spain and United Kingdom.
- Product codes and lots
- Lot Numbers: 60244767, 60254985, 60270484, 60283990, 60291761, 60295682,60298345, 60327534, 60332316, 60337990, 60347614, 60363922, 60395585, 60406637, 60416986, 60453034, 60468964, 60482887, 60517611, 60688076 and 60810866.