Z-0816-2009
King Systems 72" Universal Flex2 Extension Kit, 50 per case, King Systems Corporation, Noblesville, IN; REF #...
King Systems Corporation. / initiated 2008-08-01 / Terminated
King Systems Corp. began a recall of King Systems 72" Universal Flex2 Extension Kit, 50 per case, King Systems Corporation, Noblesville, IN; REF # XF75 on . The reason FDA records is "The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0816-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- King Systems Corporation.
- Product
- King Systems 72" Universal Flex2 Extension Kit, 50 per case, King Systems Corporation, Noblesville, IN; REF # XF75.
- Product code
- CAI Circuit, Breathing (W Connector, Adaptor, Y Piece)
- Reason for recall
- The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.
- FDA root cause
- Pending
- Action
- Consignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand.
- Quantity in commerce
- Approx. 500,000 units of all products.
- Distribution
- Worldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru.
- Product codes and lots
- Lots IKAE1, IKCJ4, IKER6, IKGC0, IKIJ7, IKJK6, IKMI3, IKQ81, IKUX4, IKVS2, IKY06, IL161, IL1Y6, IL383, IL7N9 and IL8Y9.