Z-2112-2009

EnteraLite Infinity Enteral Feeding Pump, REF INFP01, Zevex Enteral Nutrition Delivery Systems, Salt Lake...

Zevex, Inc. / initiated 2008-12-18 / Terminated / root cause: software

Zevex International, Inc. began a recall of EnteraLite Infinity Enteral Feeding Pump, REF INFP01, Zevex Enteral Nutrition Delivery Systems, Salt Lake City, UT 84123. A rotary peristaltic pump designed to... on . The reason FDA records is "Enteral feeding pumps may not accurately deliver fluid, potentially under-infusing or over-infusing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2112-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zevex, Inc.
Product
EnteraLite Infinity Enteral Feeding Pump, REF INFP01, Zevex Enteral Nutrition Delivery Systems, Salt Lake City, UT 84123. A rotary peristaltic pump designed to deliver programmed doses of enteral nutrition solutions at selectable rates.
Product code
LZH Pump, Infusion, Enteral
Reason for recall
Enteral feeding pumps may not accurately deliver fluid, potentially under-infusing or over-infusing.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
All of Zevex' customers and distributors were notified by an Urgent Medical Device Correction letter, dated 12/18/2008. The letters explained the issue, the patient impact, and what necessary actions should be taken. Customers were instructed to contact Zevex to schedule the upgrade of their installed base of pumps at 800-970-2337 or [email protected]. Distributors were to contact their customers and notify them of the letter.
Quantity in commerce
32,545
Distribution
Nationwide, including 1 VA facility in CA. Foreign distribution to 1 consignee in Canada.
Product codes and lots
Serial numbers in range: 506297001 - 508290050