Triage Drugs of Abuse Panel, Model Number: 95000. Is an immunoassay test kit used for the qualitative...
Biosite Inc began a recall of Triage Drugs of Abuse Panel, Model Number: 95000. Is an immunoassay test kit used for the qualitative determination of the presence of the major metabolites of... on . The reason FDA records is "The recall was initiated after the firm's investigations confirmed that, while lot #186625 passed all internal specifications and quality requirements, the incidence of false positive PCP results has exceeded their post...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1088-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Triage Drugs of Abuse Panel, Model Number: 95000. Is an immunoassay test kit used for the qualitative determination of the presence of the major metabolites of drugs of abuse in urine.
- Product code
- DKZ Enzyme Immunoassay, Amphetamine
- Reason for recall
- The recall was initiated after the firm's investigations confirmed that, while lot #186625 passed all internal specifications and quality requirements, the incidence of false positive PCP results has exceeded their post market surveillance thresholds for customer complaints in the field. In addition, the firm has been able to reproduce false positive results reported by customers and as a result
- FDA root cause
- Other
- Action
- Consignee notification was initiated on 11/6/2008 with Urgent Medical Device Recall notifications, dated 11/5/2008 and mailed to domestic customers sent via US Postal Service Priority Mail and to foreign customers via FedEx. The notice informed customers of the reason for the recall and requested the customers immediately discontinue all use of lot #186625 and discard all affected product in accordance with your local regulations. The customers were also instructed to complete and fax an attached Customer/Distributor Verification form within 10 days.
- Quantity in commerce
- 2634 kits
- Distribution
- Worldwide Distribution -- USA and country of Canada.
- Product codes and lots
- Lot/Unit Codes: 186625