Z-0879-2009

Biomet Microplasty Elite Instruments Vanguard Slidex Femoral Block, 70 mm, Biomet Orthopedics, Inc., Warsaw...

Zimmer GmbH / initiated 2008-12-18 / Terminated

Biomet, Inc. began a recall of Biomet Microplasty Elite Instruments Vanguard Slidex Femoral Block, 70 mm, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-485156. A manual surgical instrument... on . The reason FDA records is "Falling component: The universal pin, which holds the sliding instrument saw guide in its slot, may loosen during surgical procedures and fall into the wound". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0879-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Microplasty Elite Instruments Vanguard Slidex Femoral Block, 70 mm, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-485156. A manual surgical instrument for general use; a non-powered, hand-held, or hand-manipulated device; instrument guide.
Product code
HTZ Instrument, Cutting, Orthopedic
Reason for recall
Falling component: The universal pin, which holds the sliding instrument saw guide in its slot, may loosen during surgical procedures and fall into the wound.
FDA root cause
Other
Action
Consignees were notified by letter entitled "URGENT MEDICAL DEVICE REMOVAL NOTICE" dated 12/18/08 advising them of the problem and that the instruments should be removed from use and returned. For further information, please contact Biomet, Inc., by telephone on 800-348-9500.
Quantity in commerce
444 of all products
Distribution
Worldwide Distribution --- including USA and countries of Australia, Canada, Japan and the United Kingdom.
Product codes and lots
Lots 084400 and 132802.