Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1100-2009

Dual Port Wizard Low Profile Replacement Gastrostomy Device, REF 00204W, 20F x 1.7cm, 10cc, Sterile EO, Bard...

Bard Access Systems, Inc. / initiated 2008-12-22 / Terminated / root cause: manufacturing process

C R Bard Inc began a recall of Dual Port Wizard Low Profile Replacement Gastrostomy Device, REF 00204W, 20F x 1.7cm, 10cc, Sterile EO, Bard Access Systems, Inc., Salt Lake City, UT 84116. A... on . The reason FDA records is "Out of specification replacement gastrostomy device was distributed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1100-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
Dual Port Wizard Low Profile Replacement Gastrostomy Device, REF 00204W, 20F x 1.7cm, 10cc, Sterile EO, Bard Access Systems, Inc., Salt Lake City, UT 84116. A silicone low-profile balloon-type device is designed for percutaneous insertion through an established, appropriately sized stoma tract. It is a component of the Dual Port Wizard Gastrostomy Device.
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
Out of specification replacement gastrostomy device was distributed.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by "Product Recall" letter distributed on January 7, 2009 via Federal Express overnight mail and asked to check all inventory locations, remove unused affected product and return to Bard Access Systems, Inc (BAS). Customers were also instructed to complete the enclosed Reply Form and Inventory Reconciliation Form and fax it to BAS at 1-801-522-5674. Additional information is available by calling BAS Customer Service at 1-800-290-1689.
Quantity in commerce
1072 units
Distribution
Worldwide Distribution, USA (states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, NH, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV) and countries of Canada, Japan, Finland, Norway, Sweden, Austria, Belgium, Holland, United Kingdom, Italy, France, Switzerland, Spain and Germany, Australia, and Latin America.
Product codes and lots
Lot numbers: HUSA0401, HUSC0047, HUSC0048, HUSD0284, HUSD0329, HUSD1991, HUSD1992, HUSE0948, HUSE1558, HUSF1912, HUSG1096, HUSG1098, HUSH1634, HUSH1654, HUSI2702 and HUSI2753.