Z-1042-2009
Diasorin LIAISON (R) Chemiluminescence Analyzer Model # 15970. Colorimeter, Photometer, Spectrophotometer for...
Diasorin Inc. / initiated 2008-12-05 / Terminated / root cause: design
Diasorin Inc. began a recall of Diasorin LIAISON (R) Chemiluminescence Analyzer Model # 15970. Colorimeter, Photometer, Spectrophotometer for clinical use on . The reason FDA records is "DiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1042-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- Diasorin LIAISON (R) Chemiluminescence Analyzer Model # 15970. Colorimeter, Photometer, Spectrophotometer for clinical use.
- Product code
- JJQ Colorimeter, Photometer, Spectrophotometer For Clinical Use
- Reason for recall
- DiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- DiaSorin Inc authorized Field Service Engineer has visited the sites and performed the necessary field correction. Upon completion of their work, the field service engineer generated a Field Service Report and a copy was left with the customer.
- Quantity in commerce
- 9
- Distribution
- Worldwide Distribution -- USA including states of NC, NJ, PA, CA, WA, MO, UT, and OR and country of Canada.
- Product codes and lots
- Serial numbers: 2229000828, 2229000829, 2229000839, 2229000857, 2229000859, 2229000926, 2229000972, 2229001196, and 2229001230.