Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1013-2009

Triathlon X3 Posterior Stabilized Tibial Insert Use with Triathlon PS Femoral Components. Sze 7, THKNS 13 m...

Howmedica Osteonics Corp / initiated 2008-12-18 / Terminated / root cause: labeling and UDI

Stryker Howmedica Osteonics Corp. began a recall of Triathlon X3 Posterior Stabilized Tibial Insert Use with Triathlon PS Femoral Components. Sze 7, THKNS 13 m and SZE 8, THKNS 13mm; Catalog number 5532-G-713... on . The reason FDA records is "Stryker Orthopaedics became aware of a lot for lot mix-up between part numbers 5532-G-713, lot code LBD528 (Triathlon X3 Posterior Stabilized Tibial Insert, size 7-13 mm) and part 5532-G-813, lot code LBD512 (Triathlon...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1013-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Triathlon X3 Posterior Stabilized Tibial Insert Use with Triathlon PS Femoral Components. Sze 7, THKNS 13 m and SZE 8, THKNS 13mm; Catalog number 5532-G-713, Catalog number 5532-G-813 Stryker Orthopaedics
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Stryker Orthopaedics became aware of a lot for lot mix-up between part numbers 5532-G-713, lot code LBD528 (Triathlon X3 Posterior Stabilized Tibial Insert, size 7-13 mm) and part 5532-G-813, lot code LBD512 (Triathlon X3 Posterior Stabilized Tibial Insert Size 8-13 mm)
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Urgent Product Recall letters were sent via Federal Express on December 18, 2008 with return receipt to Stryker branches, Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. The letter stated the issue, the potential hazards, and the risk mitigation factors. The firm asked customers to examine their inventory and hospital locations to identify affected product; retrieve all affected product and return it to their branch or agency warehouse for reconciliation; reconcile all products on the attached Product Accountability Form; and return affected product. Questions should be directed to Karen Ariemma at 201-831-5718.
Quantity in commerce
9 of lot code LBD528; 14 of lot code LBD512
Distribution
Nationwide Distribution: FL and IL.
Product codes and lots
Lot code LBD528 and Lot code LBD512.