Z-0346-2009

Integris BH5000

Philips North America LLC / initiated 2008-12-10 / Terminated

Philips Medical Systems North America Co. Phillips began a recall of Integris BH5000 on . The reason FDA records is "Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0346-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Integris BH5000
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
FDA root cause
Radiation Control for Health and Safety Act
Action
An Electronic Product Radiation Warning, dated October 2008, was sent to customers. The letter identified the problem and under what circumstances that it could occur; the actions that should be taken by the customer/user to prevent risks for patients or users; and the actions planned by Philips to correct the problem. Further information or support concerning the issue could be obtained by contacting Bob Pettis at 425-487-7469 or the Philips Call Center at 800-722-9377. Firm FCO tracking numbers 72200107, 72200108, 72200111, 72200112, 72200117, 72200118, 72200119, 72200122, 72200123, 72200124, 72200125, 72200129, 72200142, 7989980, 7080085, 70600034, 70600035, 72200132, 72200133, 72200137, 72200138, 72200140 and 72200141. C & R # 1217116-10/20/08-016C
Quantity in commerce
190 units
Distribution
Worldwide Distribution
Product codes and lots
FCO 72200129 for 1st phase FCO 72200137 or 72200140 or 72200138 or 72200141 for 2nd phase Equipment #'s 33290, 76992, 103526, 103581, 103921, 104020, 104021, 104068, 104386, 104628, 104926, 104959, 105184, 105399, 105581, 105585, 105654, 504655, 504733, 504742, 504779, 504868, 504903, 504921, 505067, 505068, 505120, 505356, 505380, 505530, 505611, 505703, 506008, 506026, 506069, 506509, 506655, 519118, 519136, 519196, 519197, 519678, 519730, 519981, 520582, 520666, 520688, 520756, 520949, 520964, 521149, 521325, 521525, 522136, 522456, 530396, 530654, 530660, 531649, 531758, 531799, 531939, 532152, 532218, 532240, 532241, 532383, 533288, 534786, 535345, 535761, 535775, 537175, 537593, 538172, 549832, and 44147585.   Additional Equipment #s from Phase 2 - 6967, 10329, 13290, 13294, 13511, 13992, 13993, 14021, 17500, 17626, 25974, 25992, 26064, 35437,, 35485, 37908, 38017, 38028, 38062, 38109, 38329, 38426, 38790, 38792, 38933, 41050, 41080, 41115, 41129, 41340, 41355, 41411, 41423, 41455, 41458, 44714, 44719, 47106, 47127, 50375, 50388, 50392, 50419, 52465, 52470, 52488, 59390, 59418, 59422, 62508, 62519, 62531, 62559, 62606, 62671, 62775, 62776, 62779, 62792, 62806, 62968, 62998, 68255, 76198, 76365, 76395, 76460, 76560, 76572, 76580, 76941, 76953, 82519, 82521, 82608, 82609, 82663, 82709, 83311, 83327, 83823, 83834, 83927, 84469, 84487, 84734, 84827, 85018, 85095, 85150, 85497, 86087, 86415, 86419, 86992, 87068, 87186, 100641, 102523, 102790, 102949, 103003, 104388, 250073, 520720, 542784, 557725, 557726, 557738, 557740, 557761, 557762, 557767, 557770, 557773, 557957, 558028, 558180, 558183, 558287,49788214, 49788954, and 49794641.