Orthosoft Navitrack, OS Unicondylar Knee 1.0 - Universal; software application for orthopedic surgical...
Zimmer Inc. began a recall of Orthosoft Navitrack, OS Unicondylar Knee 1.0 - Universal; software application for orthopedic surgical stereotaxic instrument on . The reason FDA records is "Incorrect parameters: If the user switches from one surgeon profile to another surgeon profile after completing anatomical landmark acquisition tasks, the application will continue to use the initial surgeon profile...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0985-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Orthosoft Navitrack, OS Unicondylar Knee 1.0 - Universal; software application for orthopedic surgical stereotaxic instrument.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Incorrect parameters: If the user switches from one surgeon profile to another surgeon profile after completing anatomical landmark acquisition tasks, the application will continue to use the initial surgeon profile parameter values in the calculation of the predicted hip-knee-ankle alignment estimates, which may have different resection or component thickness parameters than are appropriate for
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Zimmer notified consignees via letter dated 8/19/08, which advised them not to change the selected surgeon profile after the acquisition of landmarks has begun, and the issued will be fixed with the next software release.
- Quantity in commerce
- 75 (34 U.S. and 41 international)
- Distribution
- Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Singapore, South Africa, Switzerland and United Kingdom.
- Product codes and lots
- All units with software versions/distribution numbers Orthosoft-unicondylar 1.0.2.31, Orthosoft-unicondylar 1.0.1.28 and Orthosoft-unicondylar 1.0.0.0.111.