Z-0913-2009

Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push Method, latex free, sterile, Boston Scientific...

Boston Scientific Corporation / initiated 2008-12-30 / Terminated

Boston Scientific Corp began a recall of Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push Method, latex free, sterile, Boston Scientific Corp., Natick, MA; Cat. no./REF 6653, inner pouch... on . The reason FDA records is "The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0913-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push Method, latex free, sterile, Boston Scientific Corp., Natick, MA; Cat. no./REF 6653, inner pouch UPN M00566530 and outer carton UPN M00566531. The devices are indicated for providing nutrition directly into the stomach in adult and pediatric patients who are unable to consume nutrition by conventional means.
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
FDA root cause
Other
Action
Consignees were notified by "Urgent Medical Device Recall" letter dated 12/30/08. The letter described the problem with the product, instructed customers to determine whether they store these products in the outer carton or the inner pouch at their facility, to locate all affected products, and to return them to the firm. For additional information, contact Boston Scientific Corporation at 1-508-683-4678.
Quantity in commerce
2460 of all products
Distribution
Nationwide.
Product codes and lots
Inner pouch (UPN M00566530) packaged in outer cartons (UPN M00566531) with lot number 12080466 on both pouches.