Z-0907-2009
Expressaire, Disposable Tourniquet Cuff 34", 1 line (18), sterile, Manufactured for Sammons Preston, A...
Hill-Rom Inc. / initiated 2008-11-24 / Terminated / root cause: manufacturing process
Virtus, Inc. began a recall of Expressaire, Disposable Tourniquet Cuff 34", 1 line (18), sterile, Manufactured for Sammons Preston, A Patterson Medical Company, Cedarburg, WI.; Cuff type... on . The reason FDA records is "The inflation tube may come loose at the connection to the cuff during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0907-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Expressaire, Disposable Tourniquet Cuff 34", 1 line (18), sterile, Manufactured for Sammons Preston, A Patterson Medical Company, Cedarburg, WI.; Cuff type A70107010. Each cuff is marked ExpressAire by Sammons Preston and A70107010. The tourniquet cuff is placed on a patient's thigh during an operation procedure of the lower leg. The cuff is not operational until it is attached to an air pump supply that is owned by the care giving facility (hospital).
- Product code
- KCY Tourniquet, Pneumatic
- Reason for recall
- The inflation tube may come loose at the connection to the cuff during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Sammons Preston was notified of the recall by letter dated 11/24/08, which instructed them to examine their inventory immediately for the specific lot numbers. The quantity on hand should be indicated on the table attached to the letter and the info should be faxed to Virtus at 812-933-0749. Customers that may have tourniquet cuff inventory of identified lot numbers should be contacted. Once recall items have been located in quarantine, Tara Collins, Quality Technician, should be contacted at 812-933-1121 for Return Material Authorization number.
- Quantity in commerce
- 3654
- Distribution
- Nationwide Distribution --- including state of Wisconsin.
- Product codes and lots
- Lots SO590, SO1205, and SO1300.