Z-0945-2009

IMPAX Cardiovascular Polling Procedure version 1.00.09, used in IMPAX 6.2.1 and 6.3.1.; Catalog numbers...

AGFA Healthcare Corp. / initiated 2008-12-24 / Terminated / root cause: software

AGFA Corp. began a recall of IMPAX Cardiovascular Polling Procedure version 1.00.09, used in IMPAX 6.2.1 and 6.3.1.; Catalog numbers: E18R4000, L5NQJ000, L57OP000, LBCDB000, and E2ZCK000... on . The reason FDA records is "Polling procedure may fail when multiple merge/split is carried out". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0945-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AGFA Healthcare Corp.
Product
IMPAX Cardiovascular Polling Procedure version 1.00.09, used in IMPAX 6.2.1 and 6.3.1.; Catalog numbers: E18R4000, L5NQJ000, L57OP000, LBCDB000, and E2ZCK000. IMPAX Cardiovascular is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. The IMPAX Centric Polling Procedure provides a mechanism to ensure that patient demographic and merge/split operations performed on the IMPAX database utilizing the IMPAX Service Tool are captured and applied to the IMPAX CV database, in order to keep patient updates from the IMPAX database synchronized with the IMPAX CV database.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Polling procedure may fail when multiple merge/split is carried out.
Root cause tag
software
FDA root cause
Software design
Action
All sites potentially affected by this product were notified of the issue directly via telephone. Each site was directed to observe a ten minute polling window as an interim mitigation until the normal mitigation by software Service Update is performed at the site by Agfa Support. An Urgent Safety Notice was sent via FED-EX on/about 12/24/2008 to all sites. The letter describes the potential issue and both interim and formal mitigation. Acknowledgment, via FAX-Back, that the information was received and understood was requested from the sites.
Quantity in commerce
5 units
Distribution
Nationwide Distribution --- including states of FL, IL, OR, SC, and SD.
Product codes and lots
Serial number: KQRCPF9, Equipment number: 15013696, ; Serial number: BQWBZ81, Equipment number: 10281800; Serial number: FB5C561, Equipment number: 10262174; Serial number: USE708NBP3, Equipment number: 15003839; and Serial number: GN6Z1E1, Equipment number: 15009724.