Z-0969-2009

GE Centricity PACS RA1000 Workstation (for diagnostic image analysis); GE Healthcare Integrated IT Solutions...

GE Healthcare / initiated 2009-01-12 / Terminated / root cause: software

GE Healthcare Integrated IT Solutions began a recall of GE Centricity PACS RA1000 Workstation (for diagnostic image analysis); GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity PACS... on . The reason FDA records is "If a user site created a custom default display protocol (DDP) to display the current exam versus the historical exam and the user-defined DDP fails, the system displays the exams using a GE default hanging protocol...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0969-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Centricity PACS RA1000 Workstation (for diagnostic image analysis); GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity PACS (Picture Archiving and Communication System) Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity PACS infrastructure.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
If a user site created a custom default display protocol (DDP) to display the current exam versus the historical exam and the user-defined DDP fails, the system displays the exams using a GE default hanging protocol. This may create a potential for misdiagnosis if the user does not review the date and time stamps in the title bars.
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare Integrated IT Solutions notified their Centricity PACS RA1000 Workstation software customers via letter (Urgent Medical Device Correction) dated 1/12/09 of the software anomaly resulting in a patient safety issue involving Default Display Protocols. The issue may create a potential for misdiagnosis if the user does not review the date and time stamps in the title bars. The letter provided the users with safety instructions to follow until the software is updated. Any questions were directed to the Remote Online Center, Field Engineer or the GE Customer Care Center at 800-437-1171. The accounts were requested to complete the enclosed customer reply form, supplying the workstation software or build number, and fax the completed form to 847-939-1479.
Quantity in commerce
778 units
Distribution
Nationwide, Canada, Bermuda, Mexico, Ecuador, Chile, and Brazil
Product codes and lots
All software versions