Z-1071-2009

Convoy Advanced Delivery Sheath Kit, Model Numbers: M0045660ST0, M0045661ST0, M0045662ST0, M0045663ST0...

Boston Scientific Corporation / initiated 2008-12-24 / Terminated / root cause: manufacturing process

Boston Scientific Corp began a recall of Convoy Advanced Delivery Sheath Kit, Model Numbers: M0045660ST0, M0045661ST0, M0045662ST0, M0045663ST0, M0045665ST0, M0045667ST0, M0045670ST0, M0045672ST0... on . The reason FDA records is "The marker band may detach from the sheath and a clinical failure could result in an embolism". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1071-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Convoy Advanced Delivery Sheath Kit, Model Numbers: M0045660ST0, M0045661ST0, M0045662ST0, M0045663ST0, M0045665ST0, M0045667ST0, M0045670ST0, M0045672ST0, M0045675ST0, M0045676ST0, M0045690ST0, M0045890ST0, M0045748ST0 and M0045749ST0. Manufactured by Thomas Medical Products, Malvern, PA. Product is intended for use to facilitate the intracardiac placement of interventional devices.
Product code
DYB Introducer, Catheter
Reason for recall
The marker band may detach from the sheath and a clinical failure could result in an embolism.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm initiated the recall on December 24, 2009 and notified consignees via "Urgent Company Initiated Medical Device Recall - Immediate Action Required" letters which explain the reason for the recall and request the affected product be returned to the Boston Scientific Corporation Distribution Center. An amendment letter with two additional Convoy Advanced Delivery Sheath Kit models (M0045748ST0 and M0045749ST0) was sent to affected customers on January 27, 2009. Direct questions about this recall to your local Sales Representative.
Quantity in commerce
31,340 of all units
Distribution
Worldwide Distribution.
Product codes and lots
Batch Numbers: S1721 4 S1899 6 S2103 9 S2287 6 S2497 2 S1724 9 S1903 2 S2105 6 S2287 7 S2513 9 S1732 2 S1913 1 S2109 5 S2295 6 S2522 7 S1733 8 S1913 2 S2115 2 S2295 9 S2524 0 S1735 6 S1913 3 S2118 9 S2296 4 S2536 0 S1737 1 S1926 5 S2123 4 S2297 1 S2553 7 S1737 2 S1930 1 S2135 6 S2299 4 S2553 8 S1737 4 S1956 3 S2151 1 S2299 5 S2555 1 S1757 4 S1962 0 S2156 4 S2301 4 S2557 8 S1766 6 S1962 1 S2161 5 S2305 1 S2562 9 S1766 7 S1964 4 S2166 4 S2306 9 S2566 1 S1766 8 S1970 4 S2166 5 S2307 0 S2566 4 S1792 6 S2015 1 S2166 6 S2308 9 S2567 6 S1792 S2015 S2180 S2309 and S2579.