Z-1959-2009
Stryker Reflex Hybrid variable self-drilling screw. Non Sterile. Manufactured by Stryker Spine SAS, Cestas...
Howmedica Osteonics Corp / initiated 2006-06-14 / Terminated / root cause: design
Stryker Spine began a recall of Stryker Reflex Hybrid variable self-drilling screw. Non Sterile. Manufactured by Stryker Spine SAS, Cestas, France. Distributed in USA by Howmedica Osteonics... on . The reason FDA records is "Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1959-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Reflex Hybrid variable self-drilling screw. Non Sterile. Manufactured by Stryker Spine SAS, Cestas, France. Distributed in USA by Howmedica Osteonics Corp., Mahwah, NJ 07430.
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker Spine issued an "Important Market Withdrawal" notice dated June 14, 2006. Consignees were advised to complete and return a Product Accountability Form via fax to 1-201-831-6069 prior to returning the affected product. Product return was to be performed during certain phase periods when upgrades were being shipped. For further questions, contact Stryker Spine at 1-201-831-5825.
- Quantity in commerce
- 109572 total screws, self tapping and self drilling.
- Distribution
- Nationwide Distribution -- Including states of NC, OR, AZ, TN, CA, MA, MD, IL, GA, FL, NJ, NM, KY, LA, PA, MI, UT, MO, OK, KS, OH, MN, NE, IN, MT, TX, AL, CO and WI.
- Product codes and lots
- All Product with the following Catalog Number/ Description: 48664010, Variable Angle Bone Screw Self Drilling, Diameter 4.0 x 10 mm; 48664012, diameter 4.0 x 12 mm; 48604014, diameter 4.0 x 14 mm; 48664016, diameter 4.0 x 16 mm; 48664018, diameter 4.0 x 18 mm.