Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 &...
BioSense Webster, Inc. began a recall of Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 & D-1257-02 on . The reason FDA records is "Biosense Webster initiated the recall because of the potential concern with one of the Printed Circuit Board (PCB) components used in the NaviStar RMT Catheters, which may cause the ablation catheter icon to shift on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0368-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biosense Webster Inc
- Product
- Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 & D-1257-02
- Product code
- LPB Cardiac Ablation Percutaneous Catheter
- Reason for recall
- Biosense Webster initiated the recall because of the potential concern with one of the Printed Circuit Board (PCB) components used in the NaviStar RMT Catheters, which may cause the ablation catheter icon to shift on the map viewer when RF energy is being applied. Therefore the user, attempting a linear ablation in the location displayed on the CARTO RMT screen, may ultimately apply RF energy to u
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm sent an "Urgent Medical Device Correction" letter to consignees on October 6, 2008 informing consignees. Users were advised to pay "close attention to the location of the ablation catheter icon on the map viewer when applying RF energy." If consignees suspect that they have a defective NAVISTAR RMT Catheter, they were instructed to notify Biosense Webster and to return the defective catheter for evaluation and refund/replacement. Customer were instructed to contact their local Biosense Webster representative for any questions or further information. No return reply was requested.
- Quantity in commerce
- 2559 units
- Distribution
- CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI
- Product codes and lots
- All lot numbers