Z-0949-2009

Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens...

Siemens Medical Solutions USA, Inc. / initiated 2008-12-20 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens Medical Solutions, Oncology Care System, Concord... on . The reason FDA records is "The range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0949-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens Medical Solutions, Oncology Care System, Concord, CA Medical charged-particle radiation therapy system.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly.
Root cause tag
design
FDA root cause
Device Design
Action
On 12/20/08, the firm issued letters via certified mail to direct consignees, informing them of the affected product and providing instructions on the recall. Contact Siemens Medical Solutions USA, Inc. at 1-925-246-8407 for additional information.
Quantity in commerce
43 units
Distribution
Nationwide and Worldwide
Product codes and lots
All units of Part Number: 8139789