Z-1047-2009

AxSYM CMV IgG Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; The AxSYM CMV IgG assay is a...

Abbott Medical / initiated 2009-01-19 / Terminated / root cause: manufacturing process

Abbott Laboratories began a recall of AxSYM CMV IgG Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; The AxSYM CMV IgG assay is a Microparticle Enzyme Immunoassay (MEIA) for the... on . The reason FDA records is "Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1047-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
AxSYM CMV IgG Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; The AxSYM CMV IgG assay is a Microparticle Enzyme Immunoassay (MEIA) for the semi-quantitative measurement of IgG antibodies to cytomegalovirus in human serum and plasma (EDTA, heparin or sodium citrate) as an indication of past or current infection with cytomegalovirus; list 4B47-20
Product code
LFZ Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus
Reason for recall
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Consignees were sent a Product Correction-Immediate Action Required letter on 1/19/09. The letter informed users of the reported increased frequency of probe crashes when using the affected reagent pack lots and provided instructions on what actions the laboratory can take if they have a defective reagent pack or experience a probe crash with their system. Users need to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading reagent pack on instrument; and discard any reagent packs that do not open properly. If a probe crash is observed prior to testing samples again, a probe crash recovery must be performed following the procedure in the Operations Manual. A copy of the communication needs to be provided to any laboratory that you may have forwarded affected lots to. Users were requested to return a customer reply letter acknowledging receipt and understanding of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.
Quantity in commerce
3,338 kits
Distribution
Worldwide Distribution -- USA including Puerto Rico and countries of Germany, Mexico, Hong Kong, and Japan.
Product codes and lots
lot 66577M100