Z-1247-2009

Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch Column...

Del Medical, Inc. / initiated 2009-01-12 / Terminated / root cause: design

Del Medical Systems Group began a recall of Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch Column Extensions; a radiographic x-ray tube suspension... on . The reason FDA records is "There is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1247-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Del Medical, Inc.
Product
Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch Column Extensions; a radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Del Medical Systems Group, 11550 West King Street, Franklin Park, IL 60131-1330. The device is used for X-Ray tube positioning.
Product code
IYB Tube Mount, X-Ray, Diagnostic
Reason for recall
There is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator.
Root cause tag
design
FDA root cause
Component design/selection
Action
Del Medical sent a "Field Correction Recall" letter with an acknowledgement form to the direct account dealers/assemblers on January 12, 2009 - February 4, 2009, informing them that required maintenance inspection of a retaining ring inside the column assembly may not be performed as required. If the retaining ring is not regularly inspected and replaced when worn or loose, there is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator. The accounts were requested to locate the affected systems and immediately take them out of service until correction by installing a Cable Mount Retention Retrofit Kit to eliminate the potential for the cable to become detached. Affected accounts were instructed to complete the enclosed acknowledgement form and return it via fax (847-288-7911) or mail to Del Medical. If there are any questions, please contact Del Medical Technical Support at 800-800-6006 or 847-288-7000.
Quantity in commerce
213 devices
Distribution
Worldwide Distribution -- US, Canada, Scotland and Pakistan.
Product codes and lots
All Model CTM and DCTM Ceiling Tube Mounts with either a 5 inch extension (Part 114-0304G2) or 10 inch extension (Part 114-0304G1). There are no serial numbers on the extensions.