Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 11.5 mm, 5.5 Fr, 20 mm...

Boston Scientific Corporation / initiated 2009-01-08 / Terminated / root cause: packaging

Boston Scientific Corp began a recall of Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 11.5 mm, 5.5 Fr, 20 mm, sterile, Boston Scientific Corp., Natick, MA; REF 3515... on . The reason FDA records is "The ultraviolet light barrier on the pouch may be discolored, which will allow ultraviolet light into the package and may cause the latex retrieval balloon to become brittle and to leak or burst when inflated". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1012-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 11.5 mm, 5.5 Fr, 20 mm, sterile, Boston Scientific Corp., Natick, MA; REF 3515. The device is used in diagnostic or therapeutic endoscopic retrograde cholangiopancreatography, transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, to remove stones from the biliary system, and to facilitate injection of contrast medium while occluding the duct with the balloon.
Product code
LQR Dislodger, Stone, Biliary
Reason for recall
The ultraviolet light barrier on the pouch may be discolored, which will allow ultraviolet light into the package and may cause the latex retrieval balloon to become brittle and to leak or burst when inflated.
Root cause tag
packaging
FDA root cause
Packaging
Action
Consignees were notified via letter dated January 8, 2009 to discontinue use of the affected product and to return it to Boston Scientific. Customers were instructed to complete and fax the Reply Verification Tracking Form to Boston Scientific at 508-683-5578. Direct questions about this recall to your local Sales Representative.
Quantity in commerce
294 of all products.
Distribution
Nationwide.
Product codes and lots
Lot Numbers: 11988198 and 11989069.