Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 11.5 mm, 5.5 Fr, 20 mm...
Boston Scientific Corp began a recall of Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 11.5 mm, 5.5 Fr, 20 mm, sterile, Boston Scientific Corp., Natick, MA; REF 3515... on . The reason FDA records is "The ultraviolet light barrier on the pouch may be discolored, which will allow ultraviolet light into the package and may cause the latex retrieval balloon to become brittle and to leak or burst when inflated". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1012-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 11.5 mm, 5.5 Fr, 20 mm, sterile, Boston Scientific Corp., Natick, MA; REF 3515. The device is used in diagnostic or therapeutic endoscopic retrograde cholangiopancreatography, transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, to remove stones from the biliary system, and to facilitate injection of contrast medium while occluding the duct with the balloon.
- Product code
- LQR Dislodger, Stone, Biliary
- Reason for recall
- The ultraviolet light barrier on the pouch may be discolored, which will allow ultraviolet light into the package and may cause the latex retrieval balloon to become brittle and to leak or burst when inflated.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Consignees were notified via letter dated January 8, 2009 to discontinue use of the affected product and to return it to Boston Scientific. Customers were instructed to complete and fax the Reply Verification Tracking Form to Boston Scientific at 508-683-5578. Direct questions about this recall to your local Sales Representative.
- Quantity in commerce
- 294 of all products.
- Distribution
- Nationwide.
- Product codes and lots
- Lot Numbers: 11988198 and 11989069.