AxSYM Folate Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; AxSYM Folate is an ion capture assay...
Abbott Laboratories began a recall of AxSYM Folate Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; AxSYM Folate is an ion capture assay for the quantitative determination of folate in... on . The reason FDA records is "Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1049-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- AxSYM Folate Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; AxSYM Folate is an ion capture assay for the quantitative determination of folate in human serum, plasma or red blood cells on the AxSYM System; list 3C81-20
- Product code
- CGN Acid, Folic, Radioimmunoassay
- Reason for recall
- Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Consignees were sent a Product Correction-Immediate Action Required letter on 1/19/09. The letter informed users of the reported increased frequency of probe crashes when using the affected reagent pack lots and provided instructions on what actions the laboratory can take if they have a defective reagent pack or experience a probe crash with their system. Users need to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading reagent pack on instrument; and discard any reagent packs that do not open properly. If a probe crash is observed prior to testing samples again, a probe crash recovery must be performed following the procedure in the Operations Manual. A copy of the communication needs to be provided to any laboratory that you may have forwarded affected lots to. Users were requested to return a customer reply letter acknowledging receipt and understanding of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.
- Quantity in commerce
- 1,142 kits
- Distribution
- Worldwide Distribution -- USA including Puerto Rico and countries of Germany, Mexico, Hong Kong, and Japan.
- Product codes and lots
- lot 68537M100