Z-1050-2009

AxSYM Myoglobin Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Myoglobin is a Microparticle...

Abbott Medical / initiated 2009-01-19 / Terminated / root cause: manufacturing process

Abbott Laboratories began a recall of AxSYM Myoglobin Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Myoglobin is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative... on . The reason FDA records is "Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1050-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
AxSYM Myoglobin Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Myoglobin is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative determination of myoglobin in human serum or plasma on the AxSYM System; list 3E43-20
Product code
DDR Myoglobin, Antigen, Antiserum, Control
Reason for recall
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Consignees were sent a Product Correction-Immediate Action Required letter on 1/19/09. The letter informed users of the reported increased frequency of probe crashes when using the affected reagent pack lots and provided instructions on what actions the laboratory can take if they have a defective reagent pack or experience a probe crash with their system. Users need to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading reagent pack on instrument; and discard any reagent packs that do not open properly. If a probe crash is observed prior to testing samples again, a probe crash recovery must be performed following the procedure in the Operations Manual. A copy of the communication needs to be provided to any laboratory that you may have forwarded affected lots to. Users were requested to return a customer reply letter acknowledging receipt and understanding of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.
Quantity in commerce
822 kits
Distribution
Worldwide Distribution -- USA including Puerto Rico and countries of Germany, Mexico, Hong Kong, and Japan.
Product codes and lots
lot 63493M200