Z-1202-2009

Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI. The device intended use...

Horiba Instruments Incorporated / initiated 2009-02-02 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI. The device intended use is a clinical chemistry reagent for the in vitro... on . The reason FDA records is "The product performance fails due to precipitation of the R2 component". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1202-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI. The device intended use is a clinical chemistry reagent for the in vitro, quantitative determination of direct bilirubin in serum on Hitachi analyzers.
Product code
CIG Diazo Colorimetry, Bilirubin
Reason for recall
The product performance fails due to precipitation of the R2 component.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by "Urgent: Medical Device Recall" letter dated February 2, 2009 and mailed on February 6, 2009, and instructed to conduct the recall to the retail level if they had further distributed the product, to destroy any stocks on hand, and to complete and return the "Recall Return Response Form" via mail, e-mail, or fax (734-483-1592) to Point Scientific, Inc. Direct questions to the firm's Technical Support Department at 1-800-757-5313 or 1-734-487-8300.
Quantity in commerce
24 kits
Distribution
Worldwide Distribution -- USA, Chile and Mexico.
Product codes and lots
HB936-120, Lot Number: 731801, exp. 11/2011 and HB936-294, Lot Number: 817103, exp. 5/2012.