Z-1255-2009
Roche MagNA Pure LC Workstation; Product Numbers: 12236931001, 12236931444, 12236931692 and 03670325001...
Roche Diagnostics GmbH / initiated 2009-02-09 / Terminated / root cause: design
Roche Diagnostics Corp. began a recall of Roche MagNA Pure LC Workstation; Product Numbers: 12236931001, 12236931444, 12236931692 and 03670325001. Robotic workstation for automated isolation of nucleic... on . The reason FDA records is "Protruding screw threads may cut the user when cleaning the interior of the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1255-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche MagNA Pure LC Workstation; Product Numbers: 12236931001, 12236931444, 12236931692 and 03670325001. Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material such as whole blood, serum, blood cells, culture cells, tissue, bacteria and fungi.
- Product code
- LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
- Reason for recall
- Protruding screw threads may cut the user when cleaning the interior of the instrument.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- In a letter titled "Urgent Medical Device Correction" dated February 9, 2009, consignees were notified of the protruding screws and instructed to use caution when cleaning the instrument until a protective plate is installed by the firm's field representative. Further questions should be addressed to Roche Molecular Diagnostics Technical Support at 1-800-526-1247.
- Quantity in commerce
- 371
- Distribution
- Nationwide Distribution including Puerto Rico and states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WV.
- Product codes and lots
- All units.