Z-1206-2009

Zoll AED Plus Defibrillator

ZOLL Medical Corporation / initiated 2009-02-12 / Terminated / root cause: manufacturing process

ZOLL Medical Corporation, World Wide Headquarters began a recall of Zoll AED Plus Defibrillator on . The reason FDA records is "Device fails to discharge the defibrillation energy". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1206-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZOLL Medical Corporation
Product
Zoll AED Plus Defibrillator
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Device fails to discharge the defibrillation energy.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On 3/31/09, the firm revised their recall strategy and required all users to a perform a "Mandatory Upgrade" software The software upgrade allows all potentially affected devices to monitor battery charging performance through periodic self-testing. If defective batteries are detected at any time prior to the recommended maximum of five years, users are prompted by the device to install fresh batteries. Frim issued Press Release on April 2, 2009. In the first notification Zoll notified customers by letter on 2/12/09 via certified mail. Customers instructed to replace device batteries every three years. In lieu of replacing batteries every three years, customers will also have the choice to update their devices with software that will monitor batteries for the identified defect. ZOLL will be send email notification to all consignees who have provided an email address as part of their contact information. ZOLL will also be publishing information of the notification on industry magazines. http://www.zollaedplusbatteryhelp.com/
Quantity in commerce
183,535 units
Distribution
Worldwide distribution
Product codes and lots
Serial numbers below X_ _ _200000