8 Fr 30cc Narrow Flex IAB Catheter Kit with Generica Adapters Product Number: IAB-04830-U, Intra-aortic...
Arrow International, Inc. began a recall of 8 Fr 30cc Narrow Flex IAB Catheter Kit with Generica Adapters Product Number: IAB-04830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is... on . The reason FDA records is "A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1210-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- 8 Fr 30cc Narrow Flex IAB Catheter Kit with Generica Adapters Product Number: IAB-04830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle.
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- An "Urgent Medical Device Recall" letter dated February 2, 2009 was issued to consignees via US Certified Mail to notify users of the problem. Users were instructed to check their inventory for the affected products, quarantine, and return the them to Arrow International, Inc. Customers were also instructed to follow the "Mitigation Directions" in circumstances where alternate Intra-Aortic Balloon (IAB) catheters/adaptors or other forms of therapy are not available (until receipt of available product). Canadian accounts were notified on February 9, 2009 by courier and International accounts notified on 6 February 2009 and 10 February 2009 via Registered Post/UPS. Direct questions to your local Sales Representative or the Arrow International Customer Service at 1-800-523-8446 (US) or 1-919-361-4062 (outside US). Press Release issued April 13, 2009 is posted to firm's webiste: http://www.arrowintl.com/iab_recall.asp.
- Quantity in commerce
- 1,572 units
- Distribution
- Worldwide Distribution: USA, Argentina, Australia, Taiwan, Singapore, New Zealand, India, Brazil, Belgium, Korea, Peru, Finland, and Denmark.
- Product codes and lots
- Lot/Serial Numbers: MF6079580 MF6100954 MF7033668 MF7086558 MF7086984 MF7097225 MF7097426 MF7118000 MF7118001 MF7128557 and MF8018922.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
