Z-1092-2009

GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system...

GE Healthcare / initiated 2008-03-06 / Terminated / root cause: software

GE Medical Systems, LLC began a recall of GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or... on . The reason FDA records is "GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1092-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector) Indicated for use in diagnostic and interventional angiographic procedures of human anatomy. Intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. Not intended for mammography applications.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to recover.
Root cause tag
software
FDA root cause
Software design
Action
On site visits were performed from 3/6/08 - 12/19/08 to update software on all 5 units.
Quantity in commerce
4
Distribution
Worldwide Distribution -- USA including states of MD, TX, and VA and country of China.
Product codes and lots
82416120006, 4103323100, 82416120005, and 214324INNOVA2