Z-1171-2009
Internal Sterilizable Paddles (electrodes) and Handles without discharge control for use with LIFEPAK 9 and...
Physio Control, Inc. / initiated 2009-02-16 / Terminated / root cause: labeling and UDI
Physio Control, Inc. began a recall of Internal Sterilizable Paddles (electrodes) and Handles without discharge control for use with LIFEPAK 9 and LIFEPAK 9P defibrillators/monitors on . The reason FDA records is "Sterilization efficacy can not be confirmed for the current methods of sterilization listed in the product labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1171-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Physio Control, Inc.
- Product
- Internal Sterilizable Paddles (electrodes) and Handles without discharge control for use with LIFEPAK 9 and LIFEPAK 9P defibrillators/monitors.
- Product code
- LDD Dc-Defibrillator, Low-Energy, (Including Paddles)
- Reason for recall
- Sterilization efficacy can not be confirmed for the current methods of sterilization listed in the product labeling.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Recalling firm sent Urgent Medical Device Recall letters, dated 2/2009, advising customers to discontinue use of the listed sterilization methods and to remove them from service since there is no sterilization method that can be supported by current data. They are to use the enclosed form to identify and confirm if they have the listed products and follow the instructions on the form for returning information to the firm.
- Quantity in commerce
- TOTAL: 3218 units; 2052 in the US, 1166 internationally
- Distribution
- Worldwide Distribution
- Product codes and lots
- Internal Defibrillation Handles without discharge control: Part Numbers 800441-03 thru 800441-06 Internal Defibrillation Electrodes (threaded end): Part Numbers: 802154-10 thru 802154-19 Sterilization labeling affects ALL part numbers referenced above.